- 142 -
ES CR LF
- command not recognized (incorrect mass format)
Set high checkweighing threshold HI
Format: UH_XXXXX CR LF, where: _ - space, XXXXX – mass format
Accessible responses:
UH_OK CR LF
- command carried out
ES CR LF
- command not recognized (incorrect mass format)
Give value of low checkweighing threshold HI
Format: ODH CR LF
Accessible response: DH_MASS CR LF – command carried out
Response format:
1 2 3 4-12 13 14 15 16 17 18 19
D H space mass space unit space CR LF
Mass - 9 characters with right justification
Unit - 3 characters with right justification
Give value of high checkweighing threshold LO
Format: OUH CR LF
Accessible response: UH_MASS CR LF - command carried out
Response format:
1 2 3 4-12 13 14 15 16 17 18 19
U H space mass space unit space CR LF
Mass - 9 characters with right justification
Unit - 3 characters with left justification
User log in
Format: LOGIN Name, Password CR LF (Name and password have to be
inserted as they are set in the moisture analyzer – small and capital letters)
Accessible responses:
LOGIN OK CR LF
- command understood, new user is logged in
LOGIN ERROR CR LF
- command understood, an error in name or
password occurred, log in failed
ES CR LF
- command not recognized (error in format)

Table of Contents