EasyManuals Logo

RADWAG MA 3Y User Manual

RADWAG MA 3Y
176 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #39 background imageLoading...
Page #39 background image
- 39 -
threshold, then the <Minimum sample> mode has to be enabled in the
additional setting of the weighing mode.
Procedure:
1. Press grey coloured workspace
2. The display indicates menu: Settings, Keys, Information, Printouts and
Profiles
3. Select menu <Settings> - additional options on weighing process
4. Press < Minimum sample> field
5. Below window is opened
6. Press <Method> field
7. Which opens a window containing a list of minimum sample methods
stored in balance database.
8. Select one of available entries.
9. The software returns to the previous menu window
10. Press <Mode> field
11. Which opens a window with parameter settings. Select one of below
options:
Lock on selecting this option and during the weighing process the
display indicates corresponding pictograms that inform a user whether
the weighed mass is below or above the minimum sample limit.
Additionally the software disables accepting the measurement result if it
is below the minimum sample limit
Warn on selecting this option and during the weighing process the
display indicates corresponding pictograms that inform a user whether
the weighed mass is below or above the minimum sample limit. The
user can accept a measurement that is below the minimum sample
limit, but on a printout such measurement is preceded by an asterisk
(*).
12. On setting the parameters exit the menu.
13. The mass field on the display is supplemented by an additional
informative pictogram. The pictogram changes during weighing process
and indicates current status of mass in relation to the pre-determined
value of minimum sample.