chamber of a moisture analyzer. Before starting temperature adjustment
process, place the temperature calibration kit inside the drying chamber,
following below diagrams:
Remove from the drying chamber the
following:
• disposable pan
• drying pan handle
• cross-shaped holder
• drying chamber insert
Assemble in the drying pan the
components of temperature
adjustment kit:
• drying chamber shield
• thermometer holder and the
thermometer
After assembling the kit start
temperature adjustment process.
CAUTION:
All the activities must be carried out with great caution not to damage
the measuring mechanism of the moisture analyzer.