QRAE II User’s Guide
Page 68
13 Controlled Part of Manual
The model PGM 2400 and PGM 2400P are certified according to the IECEx scheme,
ATEX and cCSAus (for U.S. and Canada). PGM 2400P is intrinsically safe and may be
used in hazardous locations.
SUBSTITUTION OF COMPONENTS MAY IMPAIR INTRINSIC SAFETY.
13.1 Marking
The product is marked with the following information:
RAE SYSTEMS
3775 N. 1
st
. St., San Jose
CA 95134, USA
TYPE PGM 2400 or PGM 2400P
Serial No/matrix: XXX-XXXX-000
IECEx CSA 07.0007X
Ex d ia IIC T4 Gb
0575
SIRA 12 ATEX 2156
II 2G Ex ia d IIC T4
Int. Safe Cl I, Dv 1
Gr. A, B, C, D, T4
C22.2 No 152-1984
ISA-12.13.01-2000
Warning: Use only 020-3402-000 or 020-3403-000 battery pack. Do not change batteries
in hazardous location. Do not mix old/new or different type of batteries. Understand the
manual before operating.
EM Immunity:
No effect when exposed to 0.43mW/cm
2
RF
interference (5-watt transmitter at 12"/10cm).
Temperature:
-20 C ≤ Tamb ≤ +50 C
Humidity:
0% to 95% relative humidity (non-condensing)
Warnings: Use only 020-3402-000 or 020-3403-000 battery pack.
Do not change batteries in hazardous locations.
Do not mix old/new or different types of batteries.
Understand the manual before operating.

Table of Contents

Other manuals for Rae QRAE II

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Rae QRAE II and is the answer not in the manual?

Rae QRAE II Specifications

General IconGeneral
BrandRae
ModelQRAE II
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals