ToxiRAE Pro User’s Guide
49
16. Alarm Signal Summary
Alarm types include:
Message/
Priority
4 beeps/flashes per second
3 beeps/flashes per second
2 beeps/flashes per second
3 beeps/flashes per second
Gas exceeds electronics’
maximum range
3 beeps/flashes per second
range
3 beeps/flashes per second
limit
3 beeps/flashes per second
limit
2 beeps/flashes per second
Gas reading measures less
than zero
Calibration failed, or needs
calibration
communcations
1 flash, 1 beep per minute plus
battery icon flashes on display
The alarm priority is as follows (in the event of multiple simultaneous alarms):
Super Alarm > Man Down Alarm > Man Down Warning > Max > Over > High > Low >
Neg > STEL > TWA > Cal Failed > Bump Failed > Cal Required > Bump Required >
Offline > Bat
* If your ToxiRAE Pro is configured with an oxygen (O
2
) sensor, alarms for TWA and
STEL are not included because neither TWA nor STEL are calculated for this sensor.