EasyManua.ls Logo

Rae ToxiRAE Pro - Operation Area and Conditions

Rae ToxiRAE Pro
68 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
ToxiRAE Pro User’s Guide
64
Operation Area and Conditions
Hazardous Areas classified by Zones
ToxiRAE Pro (PGM1860) are intended to be used in hazardous areas classified zone 0,
zone 1 or zone 2, where gases of explosion groups IIA, IIB or IIC and T4 may be present.
For North America and Canada the equipment can further be used in Class I, Zone 0.
Hazardous Areas classified by Divisions
ToxiRAE Pro is intended to be used in hazardous areas classified for Class I Div. 1 or 2,
within the temperature range of -20º C to +55º C, where gases of explosion groups A, B,
C or D and temperature code T4 may be present.
Instruction For Safe Use
Strictly follow the instructions for safe use. Application of the ToxiRAE Pro requires full
understanding and strict observation of the instructions.
Only use RAE Systems battery pack P/N: G02-3004-000
Do only charge the battery pack using RAE Systems charger outside hazardous area
(Um: 20V).
Use In Hazardous Areas
Equipment which is intended for use in explosive atmospheres and which has been
assessed and certified according to international regulations may be used only under
specified conditions. The components may not be modified in any way.
The appropriate regulations for service and repair must be properly observed during such
activities.

Table of Contents

Related product manuals