12 CLEANING
12.1 CLEANING
On page ,
OPEN
COVER LID
gives access to the inside of the machine.
Basic weekly surface cleaning is sufficient with a wet cloth.
You can use for exemple Ethanol 70°.
Clean:
- The slide drawers
- The slide holders
- The inside and outside surfaces
- The rinsing tank
Procedure of rinsing station release
Never pour liquid onto the appliance, it could damage the electronic components.
Do not use solvents on the screen and the cover lid.
In the event of accidental spillage of reagents, turn off the appliance as quickly as possible and
clean using a cloth and the appropriate solvent.
In all cases limit the dose of solvent.
12.2 HYDRAULIC SYSTEM CLEANING
This procedure should be done weekly in order to prevent the bacteria proliferation.
Equipment must be switched on.
Rinsing and draining circuits must be connected to the equipment.
Decontamination solution advised :
- Bleaching water with 2,6 % of active chlorine (available in commerce)
- Solutions like Stereanios of the Laboratoires Anios or Elusept of Pierre Fabre Santé
a. Clean the rinsing liquid container with a decontamination solution in order to prevent the proliferation of
bacteria on the container bottom.
b. Fill the rinsing liquid container with approximatively 500 ml of decontamination solution.
c. Deep the rinsing cane in the container filled with the disinfectant solution.