EasyManua.ls Logo

R&S ESU Series - Sweep Settings - SWEEP Key

R&S ESU Series
855 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
R&S ESU Instrument Functions
Analyzer Mode
4.120 Operating Manual 1302.6163.12 - 03
Fig. 4.2 Reponses of the linear and the logarithmic average detectors as a function of pulse
repetition frequency (PRF) for pulse-modulated signals with pulse durations of 400 ns
and 1 ms. Resolution bandwidth is 1 MHz.
Remote command: BAND:VID:TYPE LIN
4.8.5 Sweep Settings – SWEEP Key
The SWEEP key serves for configuring the sweep mode and opens the SWEEP
softkey menu. In split-screen mode, the entries made are valid for the active window
only.
The CONTINUOUS SWEEP, SINGLE SWEEP and SGL SWEEP DISP OFF soft-
keys are mutually exclusive selection keys.
SWEEP
CONTINUOUS
SWEEP
The CONTINUOUS SWEEP softkey activates the continuous sweep mode, which
means that the sweep takes place continuously according to the trigger mode set.
When working in the split-screen mode and with different settings in the two win-
dows, screen A is swept first, followed by screen B. When the softkey is pressed,
the sweep is restarted.
CONTINUOUS SWEEP is the default setting of R&S ESU.
Remote command: INIT:CONT ON
CONTINUOUS SWEEP
SINGLE SWEEP
CONTINUE SGL SWEEP
SWEEPTIME MANUAL
SWEEPTIME AUTO
SWEEP COUNT
SWEEP POINTS
SGL SWEEP DISP OFF

Table of Contents

Related product manuals