Operating Manual 1302.6163.12 - 03 4.297
R&S ESU Instrument Functions
Basic Settings
The FIRMWARE UPDATE softkey opens the subdirectory for installing/deinstalling
new firmware versions.
Remote command: --
FIRMWARE
UPDATE
The FIRMWARE UPDATE softkey starts the installation program and leads the user
through the remaining steps of the update.
Remote command: "SYST:FIRM:UPD 'D:\USER\FWUPDATE'"
RESTORE
FIRMWARE
The RESTORE FIRMWARE softkey restores the previous firmware version
Remote command: --
UPDATE PATH The UPDATE PATH softkey is used to select the drive and directories under which
the archive files for the firmware update are stored.
The firmware update can thus also be performed via network drives or USB memory
sticks/USB-CD-ROM drives.
Remote command: "SYST:FIRM:UPD 'D:\USER\FWUPDATE'"
APPL SETUP 
RECOVERY
The APPL SETUP RECOVERY (Application Setup Recovery) softkey controls the
instrument behaviour when changing the active application, e.g from SPECTRUM to
FM DEMOD and back from FM DEMOD to SPECTRUM.
In the default state OFF a few parameters of the current analyzer setting are passed
to the application (e.g. center frequency, level settings) or from the application back
to the analyzer mode.
If  APPL SETUP RECOVERY is switched ON, the settings of the applications are
independent of each other. Leaving the FM DEMOD application will restore the pre-
vious state of the ANALYZER. 
Remote command: SYST:APPL:SREC ON
4.9.3.9 External Noise Source 
NOISE SRC 
ON/OFF
The NOISE SRC ON/OFF softkey switches on or off the supply voltage for an exter-
nal noise source which is connected to the NOISE SOURCE connector on the rear
panel of the instrument. 
Remote command: DIAG:SERV:NSO ON
PRESET 
RECEIVER 
The RECEIVER PRESET softkey sets the receiver mode as preset mode. Default
state is the receiver mode.
Remote command: SYST:PRES:COMP OFF
PRESET 
ANALYZER
The ANALYZER PRESET softkey sets the analyzer mode as preset mode. 
The individual application settings are stored on the internal harddisk.