R&S ESU Instrument Functions
Analyzer Mode
4.130 Operating Manual 1302.6163.12 - 03
4.8.8 Selection of Trace Functions
The trace functions are subdivided as follows:
Display mode of trace (CLEAR/WRITE, VIEW and BLANK)
Evaluation of the trace as a whole (AVERAGE, MAX HOLD and MIN HOLD)
Evaluation of individual pixels of a trace using detectors
The TRACE key opens a menu offering the setting options for the selected trace.
In this menu, the mode of representing the measured data in the frequency or time
domain in the 625 pixels of the display is determined. Upon start of the measure-
ment, each trace can be displayed either completely new or based on the previous
results.
Traces can be displayed, blanked and copied. Traces can also be corrected with the
aid of mathematical functions.
The measurement detector for the individual display modes can be selected directly
by the user or set automatically by R&S ESU.
The default setting is trace 1 in the overwrite mode (CLEAR / WRITE), the other
traces 2 and 3 are switched off (BLANK).
The CLEAR/WRITE, MAX HOLD, MIN HOLD, AVERAGE, VIEW and BLANK soft-
keys are mutually exclusive selection keys.
TRACE
SELECT TRACE
CLEAR/WRITE
MAX HOLD
AVERAGE
VIEW
BLANK
SWEEP COUNT
DETECTOR
! AUTO SELECT
DETECTOR AUTO PEAK
DETECTOR MAX PEAK
DETECTOR MIN PEAK
DETECTOR SAMPLE
DETECTOR RMS
DETECTOR AVERAGE
DETECTOR QPK
DETECTOR CISPR RMS
DETECTOR CISPR AV
TRACE MATH
! T1-T2->T1 | T1-T3->T1
TRACE POSITION
TRACE MATH OFF