R&S ESU Instrument Functions
Analyzer Mode
4.140 Operating Manual 1302.6163.12 - 03
DETECTOR
The DETECTOR softkey opens a submenu for selecting the detector for the selected
trace. The softkey is highlighted if the detector is not selected with AUTO SELECT.
The detector can be selected independently for each trace. The AUTO SELECT
mode selects the optimum detector for each display mode of the trace (Clear/Write,
Max Hold or Min Hold).
The softkeys for the detectors are mutually exclusive selection keys.
AUTO SELECT The AUTO SELECT softkey (= default setting) selects the optimum detector for the
set display mode of the trace (Clear/Write, Max Hold and Min Hold) and the selected
filter mode (bandpass/FFT).
The detector activated for the specific trace is identified in the respective trace dis-
play field as follows:
Remote command: DET:AUTO ON
DETECTOR
AUTO PEAK
The DETECTOR AUTOPEAK softkey activates the autopeak detector.
Remote command: DET APE
AUTO SELECT
DETECTOR AUTO PEAK
DETECTOR MAX PEAK
DETECTOR MIN PEAK
DETECTOR SAMPLE
DETECTOR RMS
DETECTOR AVERAGE
DETECTOR QPK
DETECTOR CISPR RMS
DETECTOR CISPR AV
Trace display Detector (bandpass) Detector (FFT)
Clear/Write Auto Peak Max Peak
Average Sample Sample
Max Hold Max Peak Max Peak
Min Hold Min Peak Max Peak
Detector
Auto Peak AP
Max Peak PK
Min Peak MI
Average AV
RMS RM
Sample SA
Quasipeak QP