Operating Manual 1302.6163.12 - 03 5.25
R&S ESU Remote Control – Basics
Status Reporting System
5.7.3.6 STATus:QUEStionable:ACPLimit Register
This register comprises information about the observance of limits during adjacent
power measurements. It can be queried with commands STATus:QUEStionable:
ACPLimit:CONDition? and STATus:QUEStionable:ACPLimit[:EVENt]?.
9 LIMit (device-specific)
This bit is set if a limit value is violated (see also section “STATus:QUEStionable:LIMit<1|2>
Register” on page 5.27)
10 LMARgin
This bit is set if a margin is violated (see also section “STATus:QUEStionable:LMARgin<1|2>
Register” on page 5.28)
11 This bit is not used
12 ACPLimit
This bit is set if a limit for the adjacent channel power measurement is violated (see also
section “STATus:QUEStionable:ACPLimit Register” on page 5.25)
13 TRANSducer break
14 This bit is not used
15 This bit is always 0.
Bit
No.
Meaning
Bit
No.
Meaning
0 ADJ UPPer FAIL(Screen A)
This bit is set if in screen A. the limit is exceeded in the upper adjacent channel
1 ADJ LOWer FAIL (Screen A)
This bit is set if in screen A the limit is exceeded in the lower adjacent channel.
2 ALT1 UPPer FAIL (Screen A)
This bit is set if in screen A the limit is exceeded in the upper 1st alternate channel.
3 ALT1 LOWer FAIL (Screen A)
This bit is set if in screen A the limit is exceeded in the lower 1st alternate channel.
4 ALT2 UPPer FAIL (Screen A)
This bit is set if in screen A the limit is exceeded in the upper 2nd alternate channel.
5 ALT2 LOWer FAIL (Screen A)
This bit is set if in screen A the limit is exceeded in the lower 2nd alternate channel.
6 ALT3 to 11 LOWer/UPPer FAIL (Screen A)
This bit is set if in screen A the limit is exceeded in one of the lower or upper alternate channels
3 to 11.
7 not used
8 ADJ UPPer FAIL (Screen B)
This bit is set if in screen B the limit is exceeded in the upper adjacent channel.
9 ADJ LOWer FAIL (Screen B)
This bit is set if in screen B the limit is exceeded in the lower adjacent channel.