Operating Manual 1302.6163.12 - 03 4.185
R&S ESU Instrument Functions
Analyzer Mode
4.8.13.3 Setting the Channel Configuration
CP/ACP
CONFIG
The CP/ACP CONFIG softkey opens a submenu for configuration of the channel
power and adjacent channel power measurement independently of the offered stan-
dards.
The channel configuration includes the number of channels to be measured, the
channel bandwidths (CHANNEL BANDWIDTH), and the channel spacings (CHAN-
NEL SPACING).
Limit values can additionally be specified for the adjacent-channel power (ACP
LIMIT CHECK and EDIT ACP LIMITS) which are checked for compliance during the
measurement.
NO. OF ADJ
CHAN
The NO. OF ADJ CHAN softkey activates the entry of the number ±n of adjacent
channels to be considered in the adjacent-channel power measurement.
Numbers from 0 to 12 can be entered.
The following measurements are performed depending on the number of the chan-
nels.
NO. OF ADJ CHAN
NO. OF TX CHAN
CHANNEL BANDWIDTH
CHANNEL SPACING
ACP REF SETTINGS
CP/ACP ABS | REL
CHAN PWR / HZ
POWER MODE
! CLEAR/WRITE
MAX HOLD
ADJUST SETTINGS
Side menu
ACP LIMIT CHECK
EDIT ACP LIMITS
WEIGHTING FILTER WEIGHT TX (ON OFF)
WEIGHT TX ALPHA
WEIGHT ADJ (ON OFF)
WEIGHT ADJ ALPHA
ADJ REFLVL OFFSET
SELECT TRACE
0 Only the channel powers are measured.
1 The channel powers and the power of the upper and lower adjacent channel
are measured.
2 The channel powers, the power of the upper and lower adjacent channel and of
the next higher and lower channel (alternate channel 1) are measured.