R&S ESU Instrument Functions
Analyzer Mode
4.184 Operating Manual 1302.6163.12 - 03
SELECT USER
STD
The softkey SELECT USER STD selects and activates a user defined ACP stan-
dard.
After selecting the standard, the user defined atndard becomes available for selec-
tion in the ACP standard list (USER list item). The corresponding filename is shown
in brackets.
Remote command: CALC:MARK:FUNC:POW:PRES <file name>
SAVE AS USER
STD
As of firmware version 4.4x, you can save the current ACP configuration in a file and
use it as a user defined ACP standard. The data set is stored on the instruments
harddisk in XML file format.
Softkey SAVE AS USER STD opens a dialog box to store the user defined configu-
ration.
The following list shows the parameters that you can customize:
• Number of Adjacent Channels
• Channel Bandwidth of TX, ADJ and ALT channels
• Channel Spacings
• Resolution Bandwidth, Video Bandwidth
• ACP Limit State and ACP Limits
• Sweeptime, Sweeptime Coupling
• Detector
• Trace Mode (Clr/Write,..)
Note that the ACP User Standard is not supported for Fast ACP and Multi Carrier
ACP Measurements.
Remote command: use “SENSe:POWer Subsystem” on page 6.198 to
configure a user standard
CALC:MARK:FUNC:POW:STAN:SAVE <file name>
CALC:MARK:FUNC:POW:STAN:CAT?
DELETE USER
STD
Opens a dialog box to delete user defined ACP standards.
Remote command: CALC:MARK:FUNC:POW:STAN:DEL <standard>