EasyManua.ls Logo

R&S ESU Series - Page 168

R&S ESU Series
855 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Operating Manual 1302.6163.12 - 03 4.137
R&S ESU Instrument Functions
Analyzer Mode
Example - data section of the file
DECIM SEP
The DECIM SEP softkey selects the decimal separator between '.' (decimal point)
and ',' (comma) with floating-point numerals for the ASCII FILE EXPORT function.
With the selection of the decimal separator different language versions of evaluation
programs (e.g. Microsoft Excel) can be supported.
Remote command: FORM:DEXP:DSEP POIN
COPY TRACE
The COPY TRACE softkey copies the screen contents of the current trace into
another trace memory. The desired memory is selected by entering the number 1, 2
or 3.
Upon copying, the contents of the selected memory is overwritten and the new con-
tents displayed in view mode.
Remote command: TRAC:COPY TRACE1,TRACE2
4.8.8.1 Selection of Detector
The detectors of the R&S ESU are implemented as pure digital devices. The
following detectors are available:
“Peak detectors (MAX PEAK and MIN PEAK)” on page 4.138
“Auto peak detector” on page 4.138
“Sample detector” on page 4.138
“RMS detector” on page 4.138
“Average detector” on page 4.139
“Quasipeak detector” on page 4.139
“CISPR RMS detector” on page 4.139
“CISPR average detector” on page 4.139
File contents Description
Trace 1:;; Selected trace
x-Unit;Hz; Unit of x values:
Hz with span > 0; s with span = 0;
dBm/dB with statistics measurements
y-Unit;dBm; Unit of y values:
dB*/V/A/ depending on the selected unit with y axis LOG or % with
y-axis LIN
Preamplifier;OFF; Preamplifier state (OFF, if no preamplifier is available)
Transducer;OFF; Transducer state:
OFF:No transducer active (or available)
<trd name>selected transducer
Values; 625; Number of measurement points
10000;-10.3;-15.7
10180;-11.5;-16.9
10360;-12.0;-17.4
...;...;
Measured values:
<x value>, <y1>, <y2>
<y2> being available only with detector AUTOPEAK and containing
in this case the smallest of the two measured values for a test point.

Table of Contents

Related product manuals