R&S ESU Instrument Functions
Analyzer Mode
4.144 Operating Manual 1302.6163.12 - 03
When switched to ON (default setting), it is checked once per minute whether a fre-
quency correction for the YIG filter is required. Frequency correction is performed if
the temperature has changed by more than 3K relative to the last instance of correc-
tion.
Remote command: CORR:YIG:TEMP:AUTO ON | OFF
CAL RESULTS
The CAL RESULTS softkey calls the CALIBRATION RESULTS table, which shows
the correction data found during calibration.
The CALIBRATION RESULTS table contains the following information:
date and time of last record of correction values
overall results of correction value record
list of found correction values according to function/module
The results have the following meaning:
Remote command: CAL:RES?
The YIG CORR ON/OFF softkey is available only on models with a frequency
range 3 GHz and 40 GHz.
If the instrument is operated in a temperature-controlled environment, the cyclic
frequency drift correction can – for time-critical applications – be switched off after
an operating period of 30 minutes.
PASSED calibration successful without any restrictions
CHECK deviation of correction value larger than expected, correction
could however be performed
FAILED deviations of correction value too large, no correction was possi-
ble. The found correction data are not valid.
ABORTED calibration aborted