EasyManua.ls Logo

R&S ESU Series - Page 199

R&S ESU Series
855 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
R&S ESU Instrument Functions
Analyzer Mode
4.168 Operating Manual 1302.6163.12 - 03
Order of signals detected:
or
With 6 dB peak excursion, all signals will be detected with NEXT PEAK and NEXT
PEAK RIGHT / NEXT PEAK LEFT.
Order of signals detected:
or
AUTO MAX
PEAK
AUTO MIN
PEAK
The AUTO MAX PEAK / AUTO MIN PEAK softkeys add an automatic peak search
action for marker 1 at the end of each particular sweep. This function may be used
during adjustments of a device under test to keep track of the current peak marker
position and level.
The current marker search limit settings (LEFT LIMIT, RIGHT LIMIT, THRESHOLD,
EXCLUDE LO) are taken into account.
Remote command: CALC:MARK:MAX:AUTO ON | OFF
CALC:MARK:MIN:AUTO ON | OFF
PEAK: Signal 1
NEXT PEAK: Signal 2
NEXT PEAK: Signal 2 (no further signal detected)
PEAK: Signal 1
NEXT PEAK LEFT: Signal 1 (no further signal detected)
NEXT PEAK RIGHT: Signal 2
NEXT PEAK RIGHT: Signal 2 (no further signal detected)
PEAK: Signal 1
NEXT PEAK: Signal 2
NEXT PEAK: Signal 3
NEXT PEAK: Signal 4
PEAK: Signal 1
NEXT PEAK LEFT: Signal 3
NEXT PEAK RIGHT: Signal 1
NEXT PEAK RIGHT: Signal 2
NEXT PEAK RIGHT. Signal 4

Table of Contents