Operating Manual 1302.6163.12 - 03 4.171
R&S ESU Instrument Functions
Analyzer Mode
The TIME DOM POWER softkey activates the power measurement in the time
domain and opens a submenu for configuration of the power measurement.
The submenu allows selection of the type of power measurement (rms or mean
power), the settings for max hold and averaging as well as the definition of limits.
The power evaluation range can be limited by input of limit values.
POWER
ON | OFF
The POWER ON/OFF softkey switches the power measurement on and off. When
entering the submenu it is ON since the power measurement is already switched on
with the TIME DOM POWER softkey in the main menu.
Remote command: CALC:MARK:FUNC:SUMM:PPE ON
CALC:MARK:FUNC:SUMM:PPE:RES?
CALC:MARK:FUNC:SUMM:RMS ON
CALC:MARK:FUNC:SUMM:RMS:RES?
CALC:MARK:FUNC:SUMM:MEAN ON
CALC:MARK:FUNC:SUMM:MEAN:RES?
CALC:MARK:FUNC:SUMM:SDEV ON
CALC:MARK:FUNC:SUMM:SDEV:RES?
PEAK The PEAK softkey switches on the calculation of the peak value from the points of
the displayed trace or a segment thereof.
For the maximum peak, the largest peak value obtained since the activation of MAX
HOLD ON is displayed.
With AVERAGE ON, the peak values of a trace are averaged over several sweeps
and displayed.
The number of sweeps over which the average or the maximum value is calculated
is set with the NUMBER OF SWEEPS softkey.
Remote command: CALC:MARK:FUNC:SUMM:PPE ONCALC:MARK:
FUNC:SUMM:PPE:RES?
RMS The RMS softkey switches on the calculation of the rms value from the points of the
displayed trace or a segment of it.
For the maximum peak, the largest rms value obtained since the activation of MAX
HOLD ON is displayed.
With AVERAGE ON, the rms values of a trace are averaged over several sweeps
and displayed.
AVERAGE ON | OFF
NUMBER OF SWEEPS
This softkey is only available in time domain (span = 0).
The measurement is performed on the trace on which marker 1 is placed. To eval-
uate another trace, marker 1 should be set on another trace using the SELECT
TRACE softkey in MARKER menu (MKR).