Operating Manual 1302.6163.12 - 03 4.219
R&S ESU Instrument Functions
Analyzer Mode
DELETE
RANGE
The DELETE RANGE softkey deletes the current range. All higher ranges are set
back by one.
Remote command: LIST:RANGe<1…20>:DELete
NEXT RANGES The NEXT RANGES softkey activates the displays of the next higher subranges, i.e.
6 to 10, 11 to 15 or 16 to 20.
Remote command: --
PREVIOUS
RANGES
The PREVIOUS RANGES softkey activates the displays of the next lower sub-
ranges, i.e. 1 to 5, 6 to 10 or 11 to 15.
Remote command: --
ADJUST AXIS The ADJUST AXIS softkey automatically adjusts the frequency axis of measure-
ment diagram so that the start frequency matches the start frequency of the first
sweep range, and the stop frequency of the last sweep range.
Remote command: --
(via FREQuency:STARt <num_value> /
FREQuency:STOP <num_value>)
START MEAS Using the START MEAS softkey, you can start the measurement while the submenu
is simultaneously exited.
When the measurement is started, the R&S ESU sets up the measurement diagram
in the selected measurement window and starts the measurement in the selected
mode.
With SINGLE, a single frequency sweep occurs; afterwards the R&S ESU remains
on the stop frequency.
With CONTINUOUS, the measurement continues until it is stopped.
You can stop the measurement with STOP SWEEP.
If a STOP AFTER SWEEP point has been defined in the range, the sweep stops
automatically at the end of the respective range to allow you to change the external
circuitry, for example. This is indicated in a message box:
SWEEP Range# reached CONTINUE/BREAK
If CONTINUE is selected, the sweep is continued in the next range. If BREAK is
selected, the sweep is stopped.
Remote command: INIT:SPUR starts the measurement
INIT:CONM starts the measurement after a BREAK
has been reached
ABORt stops the measurement after a range has been
reached
STOP MEAS The STOP MEAS softkey stops the measurement. The measurement data can be
analyzed.
Remote command: ABORt