R&S ESU Instrument Functions
Analyzer Mode
4.222 Operating Manual 1302.6163.12 - 03
Table 4-1 Example: ASCII export table - file header
The data section for the measurement values starts with the key word "TRACE <n>:
", where <n> includes the number of the used trace. Next comes the peak list with
several columns also separated by ';'.
Table 4-2 Example: ASCII export table - file data section
Spreadsheet programs such as MS Excel can read this format. Use ';' as the delim-
iter for the table cells.
Remote command: MMEM:STOR:SPUR,'F:\TEST.ASC'
File contents Explanation
Type;R&S ESU; Model
Version;3.90; Firmware version
Date;02.Aug 2006; Storage date of data set
Mode;ANALYZER; SPURIOUS; Operating mode of the device
Start;9000.000000;Hz
Stop;7000000000.000000;Hz
Start/stop of the span
Unit: Hz
x-Axis;LIN; Linear (LIN) or logarithmic (LOG) (future feature)
scaling of the x-axis
Sweep Count;1; Selected number of sweeps
File contents Explanation
TRACE 1: Selected trace
Trace Mode;CLR/WRITE; Display mode of trace:
CLR/WRITE,AVERAGE,MAX HOLD,MIN HOLD,
VIEW, BLANK
x-Unit;Hz; Unit of x values:
y-Unit;dBm; Unit of y values:
Margin;6.000000:s Peak List margin
Values;8; Number of measurement points
1;1548807257.5999999000;-
65.602280;-5.602280
1;1587207214.4000001000;-
65.327530;-5.327530
1;2112006624.0000000000;-
4.388008;55.611992
Measurement values:
<Trace>;<x value>; <y value>;<delta limit>
Different language versions of evaluation programs may require different handling
of the decimal point. Using the DECIM SEP softkey, you can thus choose between
the delimiters '.' (decimal point) and ',' (comma).