Operating Manual 1302.6163.12 - 03 4.273
R&S ESU Instrument Functions
Basic Settings
Depending on the softkey selected, either the table with the data of the factor
marked (EDIT TRD FACTOR softkey) or an empty table (NEW FACTOR softkey) is
displayed. This table is empty except for the following entries:
The features of the factor are entered in the header of the table, and the frequency
and the transducer factor are entered in the columns.
During editing, a transducer factor remains stored in the background until the factor
edited is saved with the SAVE TRD FACTOR softkey or until the table is closed. A
factor that was edited by mistake can be restored by leaving the entry function.
Name - Entry of name
A maximum of 8 characters is permissible for the name. The characters have to
comply with the convention of DOS file names. If the name exceeds 8 characters,
the name is truncated. The unit automatically stores all transducer factors with the
extension .TDF. If an existing name is changed, the factor stored under the previous
name is retained and will not be overwritten automatically with the new name. The
previous factor can be deleted at a later time, using DELETE. Thus, factors can be
copied.
Remote command: CORR:TRAN:SEL <name>
Unit: dB
Interpolation: LIN for linear frequency scaling
LOG for logarithmic frequency scaling
Name Entry of name
Unit Selection of unit
Interpolation Selection of interpolation
Comment Entry of comment
FREQUENCY Entry of frequency of reference values
TDF/dB Entry of transducer factor.
ame/Unit/Interpolation: Cable dB LIN
EDIT TRAN
DU
ER FA
T
omment:
FREQUENCY FREQUENCYTDF/dB.. TDF/dB..
1.0000000 MHz
1.0000000 GHz
1.000
5.500