Operating Manual 1302.6163.12 - 03 4.277
R&S ESU Instrument Functions
Basic Settings
Remote command: CORR:TSET:UNIT <string>
Break - Activation of query when changing a subrange
The sweep can be stopped if the user changes the subrange and selects a new sub-
range of the transducer. A message informs the user that the limit has been
attained. He can continue the sweep or switch off the transducer.
The interruption is activated by setting Break to ON. Selection is by the ENTER key
which toggles between ON and OFF (toggle function).
Remote command: CORR:TSET:BRE ON|OFF
Start - Entry of start frequency of subrange
Stop - Entry of stop frequency of subrange
The individual subranges have to be linked without a gap. That is why the start fre-
quency is already defined from the second subrange (= stop frequency of previous
range).
Remote command: CORR:TSET:RANG<1.10>
<freq>,<freq>,<name>..
INS BEFORE
RANGE
The INS BEFORE RANGE softkey copies the active column and inserts it to the
right. The frequency limits can be edited.
Remote command: --
INS AFTER
RANGE
The INS AFTER RANGE softkey copies the active column and inserts it to the left.
The frequency limits can be edited.
Remote command: --
DELETE
RANGE
The DELETE RANGE softkey deletes the marked subrange. The following sub-
ranges move left.
Remote command: --
RANGES 1-5/6-
10
The RANGES 1-5/6-10 softkey switches between the display of ranges 1 to five and
ranges 6 to 10.
Remote command: --