R&S ESU Instrument Functions
Tracking Generator – Option R&S FSU-B9
4.322 Operating Manual 1302.6163.12 - 03
For information on the calibration of the reflection measurement
(CALREFLSHORT and CAL REFL OPEN) and its mechanisms, refer to section
“Calibration of Reflection Measurement” on page 4.327.
To carry out a calibration for transmission measurements the whole test setup is
through-connected (THRU).
CAL TRANS The CAL TRANS softkey triggers the calibration of the transmission measurement.
It starts a sweep that records a reference trace. This trace is then used to calculate
the difference for the normalized values.
Fig. 4.21 Calibration curve for transmission measurement
During the calibration the following message is displayed:
After the calibration the following message is displayed:
This message will be cleared automatically after approx. 3 seconds.
Remote command: CORR:METH TRAN