EasyManua.ls Logo

R&S ESU Series - Page 362

R&S ESU Series
855 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Operating Manual 1302.6163.12 - 03 4.331
R&S ESU Instrument Functions
Tracking Generator – Option R&S FSU-B9
The modulation signal is applied to the TG IN Q / FM connector.
Switching on an external FM disables the active I/Q modulation.
Remote command: SOUR:FM:STAT ON
SOUR:FM:DEV 10MHz
EXT I/Q The EXT I/Q softkey activates the external I/Q modulation of the tracking generator.
The signals for modulation are applied to the two input connectors TG IN I and
TG IN Q at the rear panel of the unit. The input voltage range is ±1 V into 50 .
Switching on an external I/Q modulation disables the following functions:
active external AM
active external FM
Functional description of the quadrature modulator:
Fig. 4.28 I/Q modulation
I/Q modulation is performed by means of the built-in quadrature modulator. The RF
signal is divided into two orthogonal I and Q components (in phase and quadrature
phase). Amplitude and phase are controlled in each path by the I and Q modulation
signal. By adding the two components an RF output signal is generated that can be
controlled in amplitude and phase.
Remote command: SOUR:DM:STAT ON
MODULATION
OFF
The MODULATION OFF softkey switches off the modulation of the tracking genera-
tor.
Remote command: SOUR:AM:STAT OFF
SOUR:FM:STAT OFF
SOUR:DM:STAT OFF
90°
I channel
Q channel
I mod
Q mod
RF IN RF OUT

Table of Contents

Related product manuals