Operating Manual 1302.6163.12 - 03 6.7
R&S ESU Remote Control – Description of Commands
Notation
<numeric_value>
<num>
These key words mark parameters which may be entered as numeric values or be
set using specific key words (character data).
The following key words given below are permitted:
– MINimum – This key word sets the parameter to the smallest possible value.
– MAXimum – This key word sets the parameter to the largest possible value.
– DEFault – This key word is used to reset the parameter to its default value.
– UP – This key word increments the parameter value.
– DOWN – This key word decrements the parameter value.
The numeric values associated to MAXimum/MINimum/DEFault can be queried by
adding the corresponding key words to the command. They must be entered fol-
lowing the quotation mark.
Example:
SENSe:FREQuency:CENTer? MAXimum
returns the maximum possible numeric value of the center frequency as result.
<arbitrary block
program data>
This key word is provided for commands the parameters of which consist of a
binary data block.