Operating Manual 1302.6163.12 - 03 6.21
R&S ESU Remote Control – Description of Commands
CALCulate Subsystem
6.5.2 CALCulate:ESPectrum Subsystem
The following chapter describes remote control commands that configure the list evaluation for the Spec-
trum Emission Mask measurement.
CALCulate<1|2>:ESPectrum:PSEarch | PEAKsearch:AUTO ON | OFF
This command activates or deactivates the list evaluation of the Spectrum Emission Mask
measurement.
Example: “CALC:ESP:PSE:AUTO OFF”
’Deactivates the list evaluation
Characteristics: RST value: ON
SCPI: device-specific
Mode: A
:CALCulate<1|2>:ESPectrum:PSEarch | PEAKsearch:MARGin -200dB … +200dB
This command sets the margin used for the limit check/peak search of the Spectrum Emission Mask
measurement.
Example: “CALC:ESP:PSE:MARG 10"
' sets the margin to 10 dB
Characteristics: *RST value: 6 dB
SCPI: device-specific
Mode: A