EasyManua.ls Logo

R&S ESU Series - Page 498

R&S ESU Series
855 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Operating Manual 1302.6163.12 - 03 6.59
R&S ESU Remote Control – Description of Commands
CALCulate Subsystem
CALCulate<1|2>:MARKer<1...4>:FUNCtion:MDEPth:RESult?
This command queries the AM modulation depth in the indicated measurement window.
A complete sweep with synchronization to sweep end must be performed between switching on the
function and querying the measured value to obtain a valid query result. This is only possible in single-
sweep mode.
The numeric suffix <1...4> of :MARKer is irrelevant for this command.
Example: "INIT:CONT OFF"
'Switches to single-sweep mode.
"CALC:MARK:X 10MHZ"
'Sets the reference marker (marker 1) to the carrier signal at 10 MHz.
"CALC:MARK:FUNC:MDEP ON"
'Switches on the modulation depth measurement in screen A.
"INIT;*WAI"
'Starts a sweep and waits for the end.
"CALC:MARK:FUNC:MDEP:RES?"
'Outputs the measured value of screen A.
Characteristics: *RST value: -
SCPI: device-specific
Mode: A
This command is only a query and therefore has no *RST value.
CALCulate<1|2>:MARKer<1...4>:FUNCtion:TOI[:STATe] ON | OFF
This command initiates the measurement of the third-order intercept point.
A two-tone signal with equal carrier levels is expected at the RF input of the instrument. Marker 1 and
marker 2 (both normal markers) are set to the maximum of the two signals. Delta marker 3 and delta
marker 4 are positioned to the intermodulation products. The delta markers can be modified separately
afterwards with the commands CALCulate:DELTamarker3:X and CALCulate:DELTamarker4:
X.
The third-order intercept is calculated from the level spacing between the normal markers and the delta
markers.
The numeric suffix <1...4> of :MARKer is irrelevant for this command.
Example: "CALC:MARK:FUNC:TOI ON"
'Switches on the measurement of the third-order intercept in screen A.
Characteristics: *RST value: OFF
SCPI: device-specific
Mode: A