R&S ESU Remote Control – Description of Commands
CALCulate Subsystem
CALCulate:STATistics:SCALe:X:RANGe 1dB to 200dB
This command defines the level range for the x-axis of the measurement diagram. The setting is
identical to the level range setting defined with the command DISPlay:WINDow:TRACe:Y:SCALe.
Example: "CALC:STAT:SCAL:X:RANG 20dB"
Characteristics: *RST value: 100dB
SCPI: device-specific
Mode: A
CALCulate:STATistics:SCALe:Y:UNIT PCT | ABS
This command toggles the scaling of y-axis between percentage and absolute.
Example: "CALC:STAT:SCAL:Y:UNIT PCT"
'toggle to percentage
Characteristics: *RST value: ABS
SCPI: device-specific
Mode: A
CALCulate:STATistics:SCALe:Y:UPPer 1E-8 to 1.0
This command defines the upper limit for the y-axis of the diagram in statistical measurements. Since
probabilities are specified on the y-axis, the entered numerical values are dimensionless.
Example: "CALC:STAT:Y:UPP 0.01"
Characteristics: *RST value: 1.0
SCPI: device-specific
Mode: A
CALCulate:STATistics:SCALe:Y:LOWer 1E-9 to 0.1
This command defines the lower limit for the y-axis of the diagram in statistical measurements. Since
probabilities are specified on the y-axis, the entered numerical values are dimensionless.
Example: "CALC:STAT:SCAL:Y:LOW 0.001"
Characteristics: *RST value: 1E-6
SCPI: device-specific
Mode: A