Operating Manual 1302.6163.12 - 03 6.125
R&S ESU Remote Control – Description of Commands
INITiate Subsystem
6.11 INITiate Subsystem
The INITiate subsystem controls the init measurement function in the selected measurement window. In
receiver mode, a distinction is made between single measurement (INITiate1) and scan (INITiate2). In
analyzer mode, a distinction is made between INITiate1 (screen A) and INITiate2 (screen B) in split
screen representation.
Fig. 6.44 State diagramm of the receiver scan
INITiate<1|2>:CONTinuous ON | OFF
This command determines whether the trigger system is continuously initiated (continuous) or
performs single measurements (single).
Setting "INITiate:CONTinuous ON" corresponds to function SCAN/SWEEP CONTinuous, i.e. the
scan/sweep of the receiver/analyzer is cyclically repeated. The setting "INITiate:CONTinuous
OFF" corresponds to function SCAN/SWEEP SINGLE.
Example: "INIT2:CONT OFF"
'Switches the sequence in screen B to single scan/sweep.
"INIT2:CONT ON"
'Switches the sequence to continuous scan/sweep.
Characteristics: *RST value: ON
SCPI: conform
Mode: R, A
INITiate<1|2>:CONMeas
In receiver mode, this command continues a stopped measurement at the current receiver frequency
in scan mode, if a scan was interrupted by a transducer stop. If the scan was interrupted with the HOLD
command, continue the scan with the command INITiate2:IMMediate.