Operating Manual 1302.6163.12 - 03 6.145
R&S ESU Remote Control – Description of Commands
MMEMory Subsystem
MMEMory:SELect[:ITEM]:ALL
This command includes all data subsets in the list device settings to be stored/loaded.
Example: "MMEM:SEL:ALL"
Characteristics: *RST value: --
SCPI: device-specific
Mode: R, A
This command is an event and therefore has no *RST value.
MMEMory:SELect[:ITEM]:CLISt ON | OFF
This command includes the scan channel list to be stored/loaded.
Example: "MMEM:SEL:CLIS ON"
'inserts the scan channel list to the list of partial data sets
Characteristics: *RST value: OFF
SCPI: device-specific
Mode: R, A
MMEMory:SELect[:ITEM]:DEFault
This command sets the default list of device settings to be stored/loaded. The latter includes:
current configuration of general device parameters (general setup)
current setting of the measurement hardware including markers
activated limit lines
user-defined color setting
configuration for the hardcopy output
activated transducers
Final measurement data
Tracking generator settings
(only in conjunction with option Tracking Generator B9 or External Generator Control B10)
Correction data for source calibration
(only in conjunction with option Tracking Generator B9 or External Generator Control B10)
Trace data, non-used transducer factors/sets and non-used limit lines are not included.
Example: "MMEM:SEL:DEF"
Characteristics: *RST value: --
SCPI: device-specific
Mode: R, A
This command is an event and therefore has no *RST value.