R&S ESU Remote Control – Description of Commands
MMEMory Subsystem
6.150 Operating Manual 1302.6163.12 - 03
File contents Explanation
File header
Type;R&S ESU; Model
Version;3.9x; Firmware version
Date;02.Feb 2007; Storage date of data set
Mode;ANALYZER;SPURIOUS; Operating mode of the device
Format for Spurious Emissions measurement
Start;9000.000000;Hz
Stop;8000000000.000000;Hz
Start/stop of the span
Unit: Hz
x-Axis;LIN; Linear (LIN) or logarithmic (LOG) (future feature) scaling of the x-
axis
Sweep Count;1; Selected number of sweeps
Range 1: Loop across all defined sweep ranges (1 to 20)
Start; 9000.000000;Hz Range start frequency in Hz
Stop; 150000.000000;Hz Range stop frequency in Hz
Filter Type;NORMAL; Range filter type: NORMAL, RRC or CFILTER
RBW;10000.000000;Hz Resolution bandwidth of the measurement filter
VBW;30000.000000;Hz Resolution bandwidth of the video filter
Auto Sweep time;ON
Sweep time; 0.145000;s Selected sweep time in the current range
Detector;RMS; Detector: MAX PEAK, MIN PEAK, RAVERAGE
REF-Level; -10.000000;dBm Setting of the reference level in the current range
Auto RF-Attenuator; OFF; Manual (OFF) or automatic (ON) setting of the RF attenuator
RF Att;15.000000;dB Range input attenuation
Sweep Points;625; Number of sweep points in the current range
Preamp; 0.000000;dB Range preamplifier, on (20 dB) or off (0 dB)
Stop after range;OFF; Stop after range, switched on (ON) or off (OFF) for the current
range
Transducer;TRD1; Transducer name (if activated)
repeated for all sweep list ranges
File data section
TRACE 1: Selected trace
x-Unit;Hz; Unit of x values:
y-Unit;dBm; Unit of y values:
Values;2500; Number of measurement points
9000.000000;-99.619965;
9225.961538;-105.416908;
9451.923077;-100.938057;
9677.884615;-99.483894;
9903.846154;-106.879539;
10129.807692;-108.772316;
Measurement values:
<x value>; <y value>