[SENSe<1|2>:]AVERage:TYPE VIDeo | LINear
This command selects the type of average function. If VIDeo is selected, the logarithmic power is
averaged and, if LINear is selected, the power values are averaged before they are converted to
logarithmic values.
The type of average calculation is equally set for all traces in one measurement window.
Example: "AVER:TYPE LIN"
'Switches screen A to linear average calculation.
Characteristics: *RST value: VIDeo