Operating Manual 1302.6163.12 - 03 6.159
R&S ESU Remote Control – Description of Commands
SENSe Subsystem
6.16.3 SENSe:CORRection Subsystem
The SENSe:CORRection subsystem controls the correction of measured results by means of frequency-
dependent correction factors (e.g. for antenna or cable attenuation).
This subsystem also controls calibration and normalization during operation with the tracking generator
options (B9/B10). The measurement windows are selected by SENSe1 (screen A) and SENSe2
(screen B).
[SENSe<1|2>:]CORRection[:STATe] ON | OFF
This command activates/deactivates the normalization of the measurement results in the selected
window provided that the tracking generator is active. The command is available only after acquisition
of a reference trace for the selected type of measurement (transmission/reflection, see command
[SENSe<1|2>:]CORRection:COLLect[:ACQire]).
Example: "CORR ON "
'Activates normalization in screen A.
Characteristics: *RST value: OFF
SCPI: conform
Mode: A
This command is only valid in conjunction with the tracking generator / ext. generator control option
(B9/B10).
[SENSe<1|2>:]CORRection:METHod TRANsmission | REFLection
This command selects the type of measurement with active tracking generator (transmission/
reflection).
Example: "CORR:METH TRAN "
'Sets the type of measurement in screen A to “transmission“.
Characteristics: *RST value: TRANsmission
SCPI: device-specific
Mode: A
This command is only valid in conjunction with the tracking generator / ext. generator control option
(B9/B10).
[SENSe<1|2>:]CORRection:COLLect[:ACQuire] THRough | OPEN
When the tracking generator is active, this command determines the type of result acquisition for the
normalization reference measurement and starts the measurement selected:
Parameter: THRough:
TRANsmission mode: calibration with direct connection between tracking
generator and device input.
REFLection mode: calibration with short circuit at the input
OPEN: only valid in REFLection mode: calibration with open input
To obtain a valid reference measurement, a complete sweep with synchronization to the end of the
sweep must have been carried out. This is only possible in the single-sweep mode.