R&S ESU Remote Control – Description of Commands
SENSe Subsystem
6.186 Operating Manual 1302.6163.12 - 03
"SENSe:LIST:POWer 935.2MHz,-20dBm,10dB,OFF,NORM,1MHz,3MHz,434us,0,
935.4MHz,-20dBm,10dB,OFF,CFIL,30kHz,100kHz,434us,0,
935.6MHz,-20dBm,10dB,OFF,CFIL,30kHz,100kHz,434us,0"
performs a measurement sequence with the following settings:
"SENSe:LIST:POWer? 935.2MHz,-20dBm,10dB,OFF,NORM,1MHz,3MHz,434us,0,
935.4MHz,-20dBm,10dB,OFF,CFIL,30kHz,100kHz,434us,0,
935.6MHz,-20dBm,10dB,OFF,CFIL,30kHz,100kHz,434us,0"
performs the same measurement and returns the result list immediately after the last frequency point.
Characteristics: *RST value: --
SCPI: device-specific
Mode: A
Step Freq.
[MHz]
Ref Level RF Att el Att Filter type RBW VBW Meas Time TRG Level
(reserved)
1 935.2 -20 dBm 10 dB OFF Normal 1 MHz 3 MHz 434 us 0
2 935.4 -20 dBm 10 dB OFF Channel 30 kHz 100 kHz 434 us 0
3 935.6 -20 dBm 10 dB OFF Channel 30 kHz 100 kHz 434 us 0
• The measurement is performed in the time domain and therefore the span is set
to 0 Hz. If the time domain is abandoned, the function is automatically switched
off.
• The measurement is not compatible with other measurements, especially as far
as marker, adjacent channel power measurement or statistics are concerned.
The corresponding commands thus automatically deactivate the function.
• The function is only available in remote control. It is deactivated when switching
the instrument back to LOCAL.