R&S ESU Remote Control – Description of Commands
SENSe Subsystem
6.186 Operating Manual 1302.6163.12 - 03
"SENSe:LIST:POWer 935.2MHz,-20dBm,10dB,OFF,NORM,1MHz,3MHz,434us,0,
935.4MHz,-20dBm,10dB,OFF,CFIL,30kHz,100kHz,434us,0,
935.6MHz,-20dBm,10dB,OFF,CFIL,30kHz,100kHz,434us,0"
performs a measurement sequence with the following settings:
"SENSe:LIST:POWer? 935.2MHz,-20dBm,10dB,OFF,NORM,1MHz,3MHz,434us,0,
935.4MHz,-20dBm,10dB,OFF,CFIL,30kHz,100kHz,434us,0,
935.6MHz,-20dBm,10dB,OFF,CFIL,30kHz,100kHz,434us,0"
performs the same measurement and returns the result list immediately after the last frequency point.
Characteristics: *RST value: --
SCPI: device-specific
Mode: A
Step Freq.
[MHz]
Ref Level RF Att el Att Filter type RBW VBW Meas Time TRG Level
(reserved)
1 935.2 -20 dBm 10 dB OFF Normal 1 MHz 3 MHz 434 us 0
2 935.4 -20 dBm 10 dB OFF Channel 30 kHz 100 kHz 434 us 0
3 935.6 -20 dBm 10 dB OFF Channel 30 kHz 100 kHz 434 us 0
The measurement is performed in the time domain and therefore the span is set
to 0 Hz. If the time domain is abandoned, the function is automatically switched
off.
The measurement is not compatible with other measurements, especially as far
as marker, adjacent channel power measurement or statistics are concerned.
The corresponding commands thus automatically deactivate the function.
The function is only available in remote control. It is deactivated when switching
the instrument back to LOCAL.