R&S ESU Remote Control – Description of Commands
SENSe Subsystem
6.206 Operating Manual 1302.6163.12 - 03
[SENSe<1|2>:]POWer:NCORrection ON | OFF
This command switches on or off the correction of the instrument inherent noise for channel power
measurement. On activating this function, a reference measurement of the instrument inherent noise
is performed. The measured noise power is then subtracted from the power in the examined channel.
The instrument inherent noise is then re-determined after any change of the center frequency,
resolution bandwidth, sweep time and level setting by repeating the reference measurement in the new
instrument setting.
Example: "POW:NCOR ON"
Characteristics: *RST value: OFF
SCPI: device-specific
Mode: A
[SENSe<1|2>:]POWer:TRACe 1 to 3
This command assigns the channel/adjacent channel power measurement to the indicated trace in the
selected measurement window. The corresponding trace must be active, i.e. its state must be different
from blank.
Example: "POW:TRAC 2"
'Assigns the measurement in screen A to trace 2.
"SENS2:POW:TRAC 3"
'Assigns the measurement in screen B to trace 3.
Characteristics: *RST value: -
SCPI: device-specific
Mode: A
The measurement of the occupied bandwidth (OBW) is performed on the trace on
which marker 1 is positioned. To evaluate another trace, marker 1 must be posi-
tioned to another trace with CALCulate:MARKer:TRACe.