Operating Manual 1302.6163.12 - 03 6.215
R&S ESU Remote Control – Description of Commands
SENSe Subsystem
Example: "SWE:EGAT ON"
'Switches on the external gate mode.
"SWE:EGAT:TYPE EDGE"
'Switches on the edge-triggered mode.
"SWE:EGAT:HOLD 100US"
'Sets the gate delay to 100 µs.
"SWE:EGAT:LEN 500US"
'Sets the gate opening time to 500 µs.
"INIT;*WAI"
'Starts a sweep and waits for its end.
Characteristics: *RST value: OFF
SCPI: device-specific
Mode: A
[SENSe<1|2>:]SWEep:EGATe:TYPE LEVel | EDGE
This command sets the type of triggering (level or edge) by the external gate signal.
The gate opening time cannot be defined with the parameter EGATe:LENGth in case of level
triggering. The gate is closed when the gate signal disappears.
Example: "SWE:EGAT:TYPE EDGE"
Characteristics: *RST value: EDGE
SCPI: device-specific
Mode: A
[SENSe<1|2>:]SWEep:EGATe:POLarity POSitive | NEGative
This command determines the polarity of the external gate signal. The setting is valid both for the edge
of an edge-triggered signal and the level of a level-triggered signal.
Example: "SWE:EGAT:POL POS"
Characteristics: *RST value: POSitive
SCPI: device-specific
Mode: A
[SENSe<1|2>:]SWEep:EGATe:HOLDoff 125 ns to 100 s
This command defines the delay time between the external gate signal and the continuation of the
sweep.
Example: "SWE:EGAT:HOLD 100us"
Characteristics: *RST value: 0s
SCPI: device-specific