Operating Manual 1302.6163.12 - 03 4.41
R&S ESU Instrument Functions
Receiver Mode
Structure of the ASCII file
Table 4-1 RECEIVER mode, final measurement data – File header
Table 4-2 RECEIVER mode, final measurement data – Data section of the file
Content of file Description
Type; R&S ESU; Instrument model
Version;2.32; Firmware version
Date;03.Aug 2004; Date record storage date
Mode;Receiver; Instrument operating mode
Start;10000;Hz
Stop;100000;Hz
Start/stop of the display range.
Unit: Hz
x-Axis;LIN; Scaling of x axis linear (LIN) or logarithmic (LOG)
Scan Count;1; Number of scans set
Transducer;TRD1; Transducer name (if switched on)
Scan 1: Loop over all defined scan ranges (1-10)
Start;150000;Hz Range - start frequency in Hz
Stop;1000000;Hz Range – stop frequency in Hz
Step;4000;Hz Range - step width in Hz for linear step width or in% (1-100)
for logarithmic step width
RBW;100000;Hz Range - resolution bandwidth
Meas Time;0.01;s Range - measurement time
Auto Ranging;ON; Auto ranging on - or off for current range
RF Att;20;dB Range - input attenuation
Auto Preamp;OFF; Auto Preamp on or off for current range
Preamp;0;dB Range - preamplifier on (20dB) or off (0dB)
Content of file Description
TRACE 1 FINAL: Selected trace
Trace Mode;CLR/WRITE; Trace mode: CLR/WRITE,AVERAGE,MAX HOLD,MIN
HOLD, VIEW, BLANK
Final Detector, QUASIPEAK; Final detector MAX PEAK, MIN PEAK, RMS, AVERAGE,
QUASIPEAK, CISPR RMS, CISPR AV
x-Unit;Hz; Unit of x values:
y-Unit;dBuV; Unit of y values:
Final Meas Time;1.000000;s Final measurement time
Margin;6.000000:s Margin
Values;8; Number of test points
2;154000.000000;81.638535;15.638
535;N;GND
1;158000.000000;86.563789;7.5637
89;N;GND
2;1018000.000000;58.689873;-
1.310127;N;GND
...
Measured values:
<Trace>;<x-value>, <y-value>; <phase>; <ground>
Phase and protective grounding are output only if a line
impedance stabilization has been activated. They specify the
setting at which the maximum RFI level at the associated
frequency was found.