Operating Manual 1302.6163.12 - 03 6.309
R&S ESU Remote Control – Description of Commands
Differences in GPIB Behavior between the FSP and the FSE Families of Instruments
FSE DISPlay[:WINDow<1|2>]:TRACe<1...4>:X:
SPACing
LINear | LOGarithmic not available for
R&S ESU
FSE DISPlay[:WINDow<1|2>]:TRACe<1...4>:X[:
SCALe]:RVALue
<numeric_value> not available for
R&S ESU
FSE DISPlay[:WINDow<1|2>]:TRACe<1...4>:X[:
SCALe]:ZOOM
ON | OFF not available for
R&S ESU
FSE DISPlay[:WINDow<1|2>]:TRACe<1...4>:X[:
SCALe]:ZOOM
[:FREQuency]:CENTer
<numeric_value> not available for
R&S ESU
FSE DISPlay[:WINDow<1|2>]:TRACe<1...4>:X[:
SCALe]:ZOOM
[:FREQuency]:STARt
<numeric_value> not available for
R&S ESU
FSE DISPlay[:WINDow<1|2>]:TRACe<1...4>:X[:
SCALe]:ZOOM
[:FREQuency]:STOP
<numeric_value> not available for
R&S ESU
R&S
ESU
DISPlay[:WINDow<1|2>]:TRACe<1...3>:Y:
SPACing
LINear | LOGarithmic |
LDB
R&S ESU:
TRACe<1...3>
LDB is not available for
FSE
FSE DISPlay[:WINDow<1|2>]:TRACe<1...4>:Y:
SPACing
LINear | LOGarithmic |
PERCent
PERCent is not available
for R&S ESU
FSE: TRACE<1...4>
R&S
ESU
FSE DISPlay[:WINDow<1|2>]:TRACe<1...3>:Y[:
SCALe]
10dB to 200dB R&S ESU:
TRACe<1...3>
FSE: TRACE<1...4>
FSET/
ESIB
DISPlay[:WINDow<1|2>]:TRACe<1...4>:Y[:
SCALe]:BOTTom
<numeric_value> not available for
R&S ESU
R&S
ESU
FSE DISPlay[:WINDow<1|2>]:TRACe<1...3>:Y[:
SCALe]:MODE
ABSolute | RELative R&S ESU:
TRACe<1...3>
FSE: TRACE<1...4>
FSE DISPlay[:WINDow<1|2>]:TRACe<1...4>:Y[:
SCALe]:PDIVision
not available for
R&S ESU
R&S
ESU
FSE DISPlay[:WINDow<1|2>]:TRACe<1...3>:Y[:
SCALe]:RLEVel
-130dBm to 30dBm R&S ESU:
TRACe<1...3>
FSE: TRACE<1...4>
R&S
ESU
FSE DISPlay[:WINDow<1|2>]:TRACe<1...3>:Y[:
SCALe]:RLEVel:OFFSet
-200dB to 200dB R&S ESU:
TRACe<1...3>
FSE: TRACE<1...4>
R&S
ESU
FSE DISPlay[:WINDow<1|2>]:TRACe<1...3>:Y[:
SCALe]:RPOSition
0 to 100 PCT R&S ESU:
TRACe<1...3>
FSE: TRACE<1...4>
R&S
ESU
FSE DISPlay[:WINDow<1|2>]:TRACe<1...3>:Y[:
SCALe]:RVALue
<numeric_value> R&S ESU:
TRACe<1...3>
FSE: TRACE<1...4>
FSE DISPlay[:WINDow<1|2>]:TRACe<1...4>:Y[:
SCALe]:RVALue:AUTO
ON | OFF not available for
R&S ESU
FSET/
ESIB
DISPlay[:WINDow<1|2>]:TRACe<1...4>:Y[:
SCALe]:TOP
<numeric_value> not available for
R&S ESU
Devices Command (Sheet 21 of 47) Parameter Notes