The ReliOn™ Micro blood glucose meter is referenced to a fresh
plasma sample. Your meter results may differ from clinical lab results.
This is due to normal variation. To compare your meter with lab results,
follow the guidelines below.
Before you compare
Perform a control solution test. This makes sure your meter and test
strips are working correctly. Wait at least 2 hours after a meal or drink
(other than water) before doing a comparison. If the lab test needs
venous blood, you must fast before giving a blood sample.
When at the lab
Blood samples for both tests must be taken and tested within
15 minutes of each other. Collect blood for the lab test with an
approved preservative. Heparin samples of blood may be used.
See test strip insert for additional information. ReliOn™ Micro
blood glucose meter is plasma referenced.
UNDERSTANDING TEST RESULTS