Troubleshooting Control Solution Test
If the control solution test result is out-of-range, check the following
and repeat the test.
Did you do the test in
controlsolutionmode?
Do you see ‘CONT’ on
thescreenwiththeresult?
Check Action
If not, warm up/cool down
to room temperature.
Were control solutions at room
temperature (72–82°F [22–28°C])
whenused?
Have the test strips and/or
controlsolutionexpired?
Does the code on the test strip bottle
matchthecodeonthemeterscreen?
If not, do the test again.
Insert a test strip; press the
button
for control solution mode.
‘CONT’ should appear on the screen.
A‘CONT’agcanalsobeattached
when results are shown on the screen.
Make sure the test strips and control
solutions are not past the expiration date.
This date is shown on bottle.
If not, obtain a new bottle of test strips.