•Triglyceridesupto3,000mg/dLdonotsignicantlyaffecttestresults.
However, glucose values in specimens beyond this triglycerides level
should be interpreted with caution.
Normal endogenous (within body) natural levels of uric acid, ascorbate
(vitamin C), bilirubin, triglycerides, and hemoglobin do not interfere with
your blood glucose results obtained.
Externally taken drugs L-dopa, dopamine, methyl-dopa, acetaminophen,
and ibuprofen will not interfere with ReliOn™ Micro blood glucose results
when taken at therapeutic concentrations.
The ReliOn™ Micro should not be used for the diagnosis of diabetes.
• Use only fresh capillary whole blood.
Do not use the ReliOn™ Micro Blood Glucose Monitoring System to test
neonates. It has not been validated for neonatal use.
•Alwaysinserttheteststripintothemeterrst,thenprickthenger.Not
doing so may result in an“E-6” appearing on the display or a test result
that is higher than the actual value.
• Do not use at altitudes higher than 10,000 ft (3,048 m).
9
Inaccurate results may occur in severely hypotensive individuals or patients
in shock. Inaccurate low results may occur for individuals experiencing
a hyperglycemic-hyperosmolar state, with or without ketosis. This meter
should not be used to test critically ill patients.