KrosFlo® TFF System – Auxiliary Scale ACSS-60K
repligen.com 420-14871-002 Rev. 00
1. Enter <P2.rEAd / 2.6.ttr> submenu.
2. Press key, filter values are displayed successively one by one: tArEH -to enter tare
value manually, tArnn – to enter tare value that is stored in scale memory.
3. Set respective value by using and keys.
4. Press key to confirm, next go to the home screen.
5.6 Tare value memory
It is possible to store 10 tare values in scale memory.
5.6.1 Entering tare value to scale memory
1. Enter <P2.rEAd / 2.7.tArn> submenu, name of tare no. 1 from tare database is displayed
(<tArE 0>), to select a different record press key.
2. Select respective entry and press key, tare value edit box is displayed.
3. Enter tare value, to do this press to select digit to be edited and to set digit value
(0 - 9).
4. Press key to confirm, <tArE 0> is displayed.
5. Press to return to home screen.
5.6.2 Selecting tare value from the scale memory
1. Enter <P2.rEAd / 2.7.tArn> submenu, name of tare no. 1 from tare database is displayed
(<tArE 0>), to select a different record press key.
2. To select a tare value press key.
3. The set tare value is displayed on the screen with a minus sign, Net symbol is shown in the
upper-left corner of the screen.
Note: The tare value acquired from the weighing device memory is not remembered upon the
weighing device restart.
5.7 Last digit
Function designed to disable display of the last weighing indication digit, this results with less
accurate measurement.
1. Enter <P2.rEAd / 2.8.LdiG> submenu.
2. Press key, filter values are displayed successively one by one, where: ALAS – Last digit
always on, nEur – Last digit always off, and uuSt – Last digit on only when the weighing
indications is stable.