EasyManua.ls Logo

REPLIGEN KrosFlo ACSS-60K

REPLIGEN KrosFlo ACSS-60K
33 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
KrosFlTFF System Auxiliary Scale ACSS-60K
User Guide
25
repligen.com 420-14871-002 Rev. 00
4. Select respective option and press key to confirm, first, text <LoAd> is displayed for 1
second, next you see a weighing window.
5. If the parts are to be weighed in a container, first put the container on a weighing pan and
tare it.
6. Load the weighing pan with the declared amount of parts. When the indication is stable (
pictogram is displayed), press key to confirm the mass.
7. Single part mass is calculated automatically, the quantity of parts (pcs) is displayed.
Note: Total weight value of all parts loaded onto the weighing pan cannot be greater than the max
capacity value.
Note: Single part mass value must be equal or greater than 0.1 of the reading unit. Unless this
condition is met, the weighing device displays a message <ErrLo>.
Note: In the course of parts quantity determination it is necessary to wait for a stable measurement
before confirming the declared quantity value.
14. Working mode - +/- control
+/- control is working mode enabling to enter checkweighing threshold values (Min, Max).
14.1 Local settings
To go to local settings, enter <3.3.HiLo> submenu.
Table 14. +/- control Local settings
Submenu Setting
3.3.1.Acc
Working mode accessibility
3.3.2.Snn
Save mode
3.3.3.Int
Time interval
3.3.4.Lo
Lo threshold
14.2 Declaring checkweighing thresholds
1. Enter <HiLo> working mode (+/- control), first, text <SEt Lo> is displayed for 1 s, next you see
a window for declaring low weighing threshold (Min).
2. Enter a respective value and press key to confirm, first, text <SEt Hi> is displayed for 1
second, next you see a window for declaring high weighing threshold (Max).
3. Enter respective value and press key for confirmation, working modes home screen is
displayed along with the declared threshold value, where: MinLoad mass lower than low
weighing thresholds, OkLoad mass within weighing thresholds, Max Load mass greater
than high weighing threshold.
Note: If the entered low threshold value (Min) is greater than the high threshold value (Max), <Err
Lo> error is displayed.