KrosFlo® TFF System – Auxiliary Scale ACSS-60K
repligen.com 420-14871-002 Rev. 00
7.1.3 Printout interval for continuous transmission
Parameter to set frequency of printout for continuous transmission. Printout interval is set in
seconds, with 0.1 [s] accuracy within 0.1 [s] – 3600 [s] range.
1. Enter <P5.1.PC / 5.1.3.Int> submenu.
2. Enter transmission rate, to do this press to select digit to be edited and to set
digit value (0 - 9).
3. Set respective value and press key to confirm, next go to the home screen.
7.1.4 Printer
<5.2.Prtr> submenu allows to:
• Select port to which the printer is connected
7.1.5 Port
1. Enter <P5.2.Prtr / 5.2.1.Prt> submenu.
2. Press key, options are displayed successively, where: nonE – No port selected, rS1 –
RS232.
3. Set respective value and press key to confirm, next go to the home screen.
7.2 Additional display
<5.3.AdSP> submenu allows to:
• Select port to which the additional display is connected
7.2.1 Port
1. Enter <P5.3.AdSP / 5.3.1.Prt> submenu.
2. Press key, options are displayed successively, where: nonE – No port selected, rS1 –
RS232 (1).
3. Set respective value and press key to confirm, next go to the home screen.
8. Printouts
8.1 Calibration report
<P6.1.CrEP> is a group of parameters allowing to declare variables that are to be printed on the
calibration report printout. Each variable allows for accessibility attribute: YES – print, no – do not
print. Calibration report is automatically generated at the end of each calibration process.