EasyManuals Logo

REPLIGEN Spectrum KR2i User Manual

REPLIGEN Spectrum KR2i
77 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #25 background imageLoading...
Page #25 background image
25 400-12355-000 Rev. 03
KR2i / KMPi TFF Systems
Setup and Operation
3. Pressure/Valve
Pressure Unit: Either psi or bar
NOTE: Green highlighted boxes are currently selected.
Valve: Enter both Valve 1 and Valve 2 (for KMPi) settings.
Valve Mode: Select to control Valve Manually or Automatically (see section 10.3.3).
Tubing Size: Select the tubing size that goes through the backpressure valve.
Control: Select the pressure control location (Feed, Permeate, Retentate, or TMP).
Starting Position: Select the starting position for automated control from following options:
1. Open no pinching of tubing
2. Half 50% closed based on selected tubing size in valve
3. Custom pinches the tubing based on the Custom Start Position value
Valve Size: select either the KR2i Valve (tubing up to 5/8” OD) or the KMPi Valve
(tubing up to ¾” OD)

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the REPLIGEN Spectrum KR2i and is the answer not in the manual?

REPLIGEN Spectrum KR2i Specifications

General IconGeneral
BrandREPLIGEN
ModelSpectrum KR2i
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals