EasyManua.ls Logo

REPLIGEN Spectrum KR2i - System Functionality Overview; Pump Controls

REPLIGEN Spectrum KR2i
77 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
35 400-12355-000 Rev. 03
KR2i / KMPi TFF Systems
Setup and Operation
7.10.4 System Settings
a. Pump Mode – Depending on user’s Auxiliary Components, user may select one of many
Pump Modes for the TFF system to operate in that allow for manual, semi-automated, and
automated processes
b. Pressure Calibrate – Using a calibrated pressure source, the user can calibrate pressure
transducer readings
c. Alarm and Stops – Safety pressure, weight and UV alarms and interlocks
d. ABV Settings - TFF system's valve settings must be configured prior to using the valves (see
Section 10.3.3 for details)
e. Tubing Calibrate – After running the tubing in the pump for ten minutes, the user may
calibrate the tubing to ensure accurate fluid dispensing; tubing can not be calibrated if pump
is set to display RPM
f. Aux Pumps – TFF system’s Aux Pump settings must be configured prior to using the Auxiliary
Pumps to ensure the TFF system sends the correct commands to each pump
g. Hardware Setup – Time, date, unit of measure, and some Pump Mode options may be set
from the Hardware Setup menu described in more detail in the next section
h. Diagnostics – Diagnostics are for the factory use only, but may be viewed by the user
7.10.5 Hardware Setup
a. Auto Tare – When “On,” TFF system will tare pressure transducer readings
b. Pressure Units – User is able to choose “psi” or “bar” (selected unit of measure is also
displayed in KF Comm software)
c. Time – Change system time
d. Date – Change system date
e. Key Beep – When “On,” the TFF system will make an audible noise when a soft button is
pressed
f. Display Contrast – Raise or lower the screen contrast

Table of Contents