KR2i / KMPi TFF Systems
Setup and Operation
g. PSI DecimalsFor PSI unit of measure onlyselect preferred number of PSI decimals (range
of 0, 1, and 2)
h. Bar DecimalsFor BAR unit of measure onlyselect preferred number of BAR decimals
(range of 1, 2, and 3)
i. Key TimeoutWhen “On,” if the TFF system’s soft buttons are untouched for ten minutes
while on any screen other than the Main Screen, the TFF system will automatically switch
back to the Main Screen
j. Running ChangeWhen “On,” the tubing size and pump direction can be changed while the
pump is running
k. No. of HeadsInput number of attached pump heads
l. Perm Stop First - Permeate Pump motor stops before main drive stops
m. Check Weight - Pump Mode will not start if Feed Scale Weight is less than Initial
Concentration Weight, or if the Permeate Scale Weight is more than the Initial Diafiltration
Weight
n. Scale Req 4 Man - Manual Mode will not start unless both Feed and Permeate Scales are
present
o. Slow for Valve - If Automatic Backpressure Valve is unable to control pressure, then main
drive will slow down gradually to control pressure increases especially when concentrating
viscous solutions
p. Ramp Seconds - Determines rate at which TFF system will increase main pump flow rate
linearly over desired amount of time
q. Minus ModesEnables TFF system to use weight set-points instead of CF/DV setpoints.
Once turned on KF Comm needs to be closed and reopened to have access to these modes
in the Pump Control Window.
r. Valve 2 - (on KMPi only) Select either 'Valve 2 Port' if using a secondary automatic
backpressure valve or 'KONDUiT' if using the conductivity, temperature and UV sensors
NOTE: when connecting a KONDUiT to the Valve 2 serial connection use the null serial connector
included with the KONDUiT.
s. Shear ControlWhen set to Off, after initial ramp rate the pump drive will run continuously
at set flow rate. When set to On, TFF system will increase main pump flow rate depending
on feedback from Permeate Scale to make up for permeated volume (e.g. Permeate Scale
read weights as increasing at ~200g/min, main pump flow rate will increase by 200ml/min)
t. Cond 1 K and Cond 2 K - enter the K factor printed on the conductivity sensor when using
the KONDUiT

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the REPLIGEN Spectrum KR2i and is the answer not in the manual?

REPLIGEN Spectrum KR2i Specifications

General IconGeneral
BrandREPLIGEN
ModelSpectrum KR2i
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals