Page #22 background imageLoading...
Page #22 background image
Using the Astral device
18
Information menu
The Information menu is comprised of three sub-menus:
Eventsall logged event activity that has taken place is displayed. A breakdown of specific alarms,
settings or system events can also be viewed.
Deviceinformation about the actual device is displayed, eg, Model and Serial numbers, software
versions, internal battery charge along with the hours since the last service and number of hours the
device has been used.
Batteryinformation about the state of charge of the internal and external batteries when connected
including the combined total battery charge.
Device settings
The configurable settings are described in the following table.
Device setting Description
Alert Tone Sets alert tones to on or off.
Default: On
Alarm Volume Sets the volume level of the alarm system.
Settings from 1, 2, 3, 4 or 5.
Default: 3
Auto power off Automatically powers off the device after 15 minutes of inactivity.
Conditions: The device is in Ventilation standby mode (not ventilating), is being powered by the Internal
battery or an External battery and there are no active alarms.
Default: On
Display Brightness Sets the brightness of the screen from Auto with a selection of five different brightness levels.
Default: Auto
Backlight timeout
Allows the screen backlight to turn off (go black) if the screen has not been touched for two minutes or
more and there are no active alarms.
Setting to 'Off' will mean the screen back-light will be permanently on.
Default: On
Rotate Display Flips the current orientation of the display.
Device Vent LED Sets the status of the Ventilation active LED to On or Off during ventilation.
Default: On

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the ResMed Astral 100 Series and is the answer not in the manual?

ResMed Astral 100 Series Specifications

General IconGeneral
Device TypePortable Ventilator
Max Pressure60 cm H2O
Operating Temperature5°C to 40°C (41°F to 104°F)
Operating Humidity10% to 95% relative humidity, non-condensing
DisplayColor LCD touchscreen
Modes of VentilationCPAP
Data StorageSD card
AlarmsAdjustable visual and audible alarms
Tidal Volume50 to 2500 mL
Inspiratory Time0.3 – 5 seconds
FiO2 MonitoringYes
Power SupplyAC 100 – 240V, 50/60 Hz
Oxygen CompatibilityYes
Pressure Range4 – 60 cm H2O
PEEP4 to 25 cm H2O