EasyManuals Logo

ResMed Astral 100 Series User Manual

ResMed Astral 100 Series
85 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #34 background imageLoading...
Page #34 background image
Assembling patient circuits
30
Oxygen Sensor Results
Icon Description
Oxygen sensor calibration has passed.
Oxygen sensor not tested or not installed.
1. If your device was supplied without an oxygen sensor, ignore this message and proceed with therapy.
2. If possible, check that the oxygen sensor is securely attached as described in Replacing the oxygen sensor.
3. Repeat Learn Circuit. If the oxygen sensor is still not detected, return the device for servicing by an authorised
ResMed Service Centre.
Oxygen sensor calibration has failed.
Below are general steps to resolve the oxygen sensor calibration issue. Refer to Learn Circuit Troubleshooting (see
page 62) for suggested actions on the error code.
1. If possible, replace the oxygen sensor as described in Replacing the oxygen sensor.
2. Repeat Learn Circuit. If problem persists, return the device for servicing by an authorised ResMed Service Centre.
If you choose to proceed with ventilation, FiO
2
alarms will be disabled. An alternate method for monitoring FiO
2
is
required.
Expiratory Flow Sensor Results
Icon Description
Expiratory flow sensor calibration has passed.
Expiratory flow sensor not tested or not installed.
Expiratory flow sensor calibration has failed.
Below are general steps to resolve the expiratory flow sensor calibration issue. Refer to Learn Circuit Troubleshooting
(see page 62) for suggested actions on the error code.
1. Remove the adapter, seal, and expiratory flow sensor.
2. Inspect the module, seal, and flow sensor for any foreign materials.
3. Re-install the module and flow sensor, ensuring that it is securely in place.
4. If possible, replace the expiratory flow sensor as described in Replacing the expiratory flow sensor.
5. Repeat Learn Circuit. If problem persists, return the device for servicing by an authorised ResMed Service Centre.
If you choose to proceed with ventilation, check Vte and MVe alarms are effective.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the ResMed Astral 100 Series and is the answer not in the manual?

ResMed Astral 100 Series Specifications

General IconGeneral
Device TypePortable Ventilator
Max Pressure60 cm H2O
Operating Temperature5°C to 40°C (41°F to 104°F)
Operating Humidity10% to 95% relative humidity, non-condensing
DisplayColor LCD touchscreen
Modes of VentilationCPAP
Data StorageSD card
AlarmsAdjustable visual and audible alarms
Tidal Volume50 to 2500 mL
Inspiratory Time0.3 – 5 seconds
FiO2 MonitoringYes
Power SupplyAC 100 – 240V, 50/60 Hz
Oxygen CompatibilityYes
Pressure Range4 – 60 cm H2O
PEEP4 to 25 cm H2O

Related product manuals