4) Following the operator’s manual for the calibration bath (or
internal procedures):
a. Insert the temperature probes into the calibration bath.
Caution: Do not submerge the probe to the point where the
interface between cable and sheath goes below the liquid surface.
Doing so could damage the probes.
b. Set the bath temperature to a setpoint below 0 °C (-20 °C is
recommended).