Calibration
CGS-240 Manual 6/15/2021
Page 49 of 148
4) Following the operator’s manual for the calibration bath (or
internal procedures):
a. Insert the temperature probes into the calibration bath.
Caution: Do not submerge the probe to the point where the
interface between cable and sheath goes below the liquid surface.
Doing so could damage the probes.
b. Set the bath temperature to a setpoint below 0 °C (-20 °C is
recommended).
5) Open Cal Mode by selecting Utilities from the drop-
down menu.
6) Click the Calibration button.